Whenever stability test are of prime importance under ICH guidelines within Q1A22 parameter BioGene Stability chamber guide through the Harmonization process to enable qualify for License. Living cells study, New Drug Development and pharmaceutical compounds, License are obtain by using BioGene Humidity chamber under the ICH-Q1AR2 Guidelines with ease.
Bio Gene test chamber are provided with state of art 21 CFR part 11 Software which makes the user comfortable.
BIOGENE Stability Chambers have found different application shelf life testing, accelerated aging, steady state in electronic components, Sim card, industrial and biological materials.
BIOGENE Stability Chambers meet Photo stability studies in Pharma, Plastics, composites, Adhesives, Textiles. Cosmetic, Food & Beverages, Personal care Biomedical storage, life science, Electronics etc in compliance with ICH guideline Q1A along with state of art 21 CFR Part 11 software.
Bio-Gene Temperature and Humidity Chamber have:
Programmable, 999 cycle large capacity program controller, various alarm function, self-diagnostic, high accuracy with 0.1%,
Built-In RS-485 Interface enables the chamber easily hook-up to personal computer
Silent Dual Turbo Fan motor circulate air and maintain uniforms temperature in the chamber and quiet operation and auto-defrost function
Safety and Convenience over temperature and current protection capability ensure user safety and adjustable sliding rack
Sliding Water Storage Tank and automatic water supplying device.
DESCRIPTION | BTI-HC-171 | BTI-HC-280 | BTI-HC-420 | BTI-HC-560 | BTI-HC-840 |
Temperature Range | 5°C to 60°C | ||||
Temperature Accuracy | ± 0.2°C | ||||
Temperature Uniformity | ± 1°C | ||||
Humidity Accuracy | ± 2% RH | ||||
Humidity Uniformity | ± 3% RH | ||||
Temperature sensor | 3 wire RTD | ||||
Humidity range | 5% RH to 90% RH | ||||
Humidity Stability | ±2% or Better | ||||
Humidity Sensor | Capacitive | ||||
Light | 15000 LUX (optional) | ||||
Interior construction | Stainless steel 304 | ||||
Exterior Construction | CRCA Powder coated | ||||
Internal Door | Glass Sheet door with complete visibility of components. | ||||
Option for electrical interface | Standard connector | ||||
Trays | SS 304 perforated. | ||||
Electrical | 220 to 240 volts AC 50 Hz single phase | ||||
Timer cyclic | 0 to 24 hrs day night | ||||
Alarm | Audio and visual | ||||
Dimensions | 505 x 415 x 830 | 570 x 550 x 875 | 700 x 640 x 900 | 700 x 700 x 1100 | 825 x 825 x 120 |
Capacity | 171 ltrs | 280 ltrs | 420 ltrs | 560 ltrs | 840 ltrs |